Install the temperature-calibration set acc. to fig.
Switch-on the heating top on the rear side.
Close cover of the heating top
Switch-on the temperature calibration set and carry out the following adjustments,
see chap. 2.2. Select unit °C, readability 1°C
Invoke service function:
In weighing mode press the MENU button, the menu item „Therm" will be dis-
played.
Confirm by pressing the PRINT button. Press repeatedly the MENU button until
„Service" will be displayed.
Keep pressed the PRINT button for 2 sec. Select the desired settings by pressing
the MENU button
DLB_A01BA-def-1110
Temperature calibration
Temperature adjustment
Settings of lamp
13