I-PRESS 4 is intended for use in the electromagnetic environment specified below. The
customer or the user of I-PRESS 4 should assure that is used in such environment.
Immunity test
Electrostatic
discharge (ESD)
EN 61000-4-2
Electrical
fast
transient/burst
IEC 61000-4-4
Impulses
EN 61000-4-5
Voltage dips, short
interruptions
and
voltage variations on
power supply input
lines
IEC 61000-4-11
Mains
power
electromagnetic field
EN 61000-4-8
Guidance and MANUFACTURER'S declaration – ELECTROMAGNETIC IMMUNITY –
FOR ALL THE EQUIPMENT AND THE SYSTEMS
I-PRESS 4 is intended for use in the electromagnetic environment specified below. The
customer or the user of I-PRESS 4 should assure that is used in such environment
IACER SRL
Test
level
EN
60601-1-2
± 6kV contact
± 8kV air
± 2kV for power
supply lines
±1kV
differential
mode
< 5% U
T
(>95% dips of U
)
T
per 0,5 cycles
40% U
T
(60% dips of U
)
T
per 5 cycles
70% U
T
(30% dips of U
)
T
per 25 cycles
< 5% U
T
(>95% dips of U
)
T
per 5 seconds
3 A/m
SUPPORTING
Page 25 de 28
Compliance level
± 6kV contact
± 8kV air
± 2kV per power
supply lines
±1kV
differential
mode
< 5% U
T
(>95% dips of U
)
T
per 0,5 cycles
40% U
T
(60% dips of U
)
T
per 5 cycles
70% U
T
(30% dips of U
)
T
per 25 cycles
< 5% U
T
(>95% dips of U
)
T
per 5 seconds
3 A/m
THAT ARE NOT LIFE-
Electromagnetic
environment
guidance
Floors
should
be
wood, concrete or
ceramic tile. If floor
are
covered
with
synthetic
material,
the relative humidity
should be at least
30%
Mains power quality
should be at that of
a typical commercial
or
hospital
environment.
Mains power quality
should be at that of
a typical commercial
or
hospital
environment.
Mains power quality
should be at that of
a typical commercial
or
hospital
environment.
If the user of the I-
PRESS 4 requires
continued operation
during power mains
interruptions,
it
is
recommended
that
I-PRESS
4
be
powered
from
an
uninterruptible
power supply or a
battery.
Mains power quality
should be at that of
a typical commercial
or
hospital
environment.
MNPG221-01 del 05/09/16