Power Connection; Input Power Connector (24 Vdc Type) 969-8995; Input Power Connector 969-8996; Controller Connection - Varian Turbo-V 81-AG Notice De Mode D'emploi

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 11

POWER CONNECTION

Input power connector (24 Vdc type) 969-8995

The following figure shows the input power connector.
Mains 24V

Input power connector 969-8996

The following figure shows the input power connector.

CONTROLLER CONNECTION

P2 – Fan
The 24 Vdc, 100 mA maximum output voltage is present after
Switch ON of the controller.
This
is
a
dedicated
Turbo-V 81-AG pump cooling fan (see orderable parts table).
If a fan is already installed, use the available adapter cable to
connect to the pump.
P1 – Vent
This is a dedicated 24 Vdc connector for control of the optional
vent valve (see orderable parts table).
The vent valve can be driven by the controller automatically or
via a serial line.
If a 110 Vac-powered vent valve has already been installed,
this valve must be replaced with a new model (see orderable
parts table).
39/50
TECHNICAL INFORMATION
output
for
the
P2 Fan Connector
P1 Vent Connector
87-901-004-01 (B)
optional

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

969-8995969-8996

Table des Matières