ergo vac & ergo vac
10.1
Guidance and manufacturer's declaration – electromagnetic immunity
The ergo vac is intended for use in the electromagnetic environment specified below.
The customer or the user of the ergo vac hould assure that it is used in such an environment.
Immunity test
standard
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical fast
transient / burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short interruptions and
voltage variations on power supply lines
IEC 61000-4-11
Note: U
is the a.c. mains voltage prior to application of the test level.
T
Power frequency (50 / 60 Hz)
magnetic field
IEC 61000-4-8
page 38
Mode d'emploi
LT
Tableaux 1:
Immunity (all devices): electromagnetic emissions
IEC 60601
test level
± 8 kV contact
±15 kV air
± 2 kV for power supply
lines
± 1 kV for
input/output lines
± 1 kV line to line
± 2 kV line to earth
<5% U
(0,5 cycle)
T
40% U
(5 cycles)
T
70% U
(25 cycles)
T
<5% U
for 5 s
T
3 A / m
Compliance
level
environment - guidance
Floors should be wood, concrete or ceramic
± 8 kV contact
tile. If floors are covered with synthetic
material, the relative humidity should be at
±15 kV air
least 30%
± 2 kV for power supply
Mains power quality should be that of a ty-
lines
pical commercial or hospital environment.
± 1 kV line to line
Mains power quality should be that of a ty-
pical commercial or hospital environment.
± 2 kV line to earth
<5% U
(0,5 cycle)
T
40% U
(5 cycles)
Mains power quality should be that of a ty-
T
pical commercial or hospital environment.
70% U
(25 cycles)
T
<5% U
for 5 s
T
200 A / m
Power frequency magnetic fields should be
at levels characteristic of a typical location
in a typical commercial or hospital environ-
ment.
Electromagnetic