when
Breakdown
• UtoU10 appears on
the display.
During
• Constant soft noise. • Such noise is normally caused by blows
weighing
• No data can be sent
or received.
• CAL E2 is
displayed.
During
adjust-
• CAL E3 is displayed • The reference weight used during
ment
• CAL E4 is displayed • The balance is defective.
8.1
Error display
If the cause is marked with an [S], contact your nearest KERN technical service
agent.
Error
Possible causes
message
CAL E2
The zero points deviates too
much from the adjustment.
CAL E3
Large deviation with respect to
PCAL.
CAL E4
The sensitivity deviates too
much from the adjustment.
CHE x
The balance stops on this
display.
Err 0x
Abnormal internal balance
chamber.
Err 20
A wrong value has been
entered.
Err21
The necessary conditions and
numeric value have not been
met.
Err24
The balance does not store the
settings properly, voltage
fluctuations.
48
Possible causes
• Command code allowed for modification
-> Unplug the power cord, then plug it
back in after 10 sec. If the balance
displays this for more than 24 hours,
accurate measurement will not be
possible.
when loading the balance. This noise is
completely normal.
• Erroneous report parameter
• There was a weight on the balance
plate during adjustment.
-> Unload the balance and readjust.
adjustment is wrong.
-> [S]
Solution
Empty the balance plate.
Use a proper weight.
Use a proper weight.
->[S]
-> [S]
Repetition of numeric value.
Check the analog mode g.
Check the power supply.
ABS/ABJ-BA-e-0916