Télécharger Imprimer la page

KERN CXB Serie Mode D'emploi page 68

Masquer les pouces Voir aussi pour CXB Serie:

Publicité

8.2 Numeric entering of the reference weight
If you know the reference weight/piece you can enter this via number keys.
Enter reference weight via number keys
a
PRE-
TARE
Confirm with
key
a
PRE-
TARE
Now you can place the parts to be counted onto the weighing plate. All quantity pa-
rameters of your goods to be weighed are displayed:
8.3 Automatic reference optimization
If it was impossible to determine a reference due to instable goods to be weighed or
an insufficient reference weight, the [
window during reference calculation.
Overlay
indicates:
Add additional parts until the [
An audio signal indicates that reference optimization has been carried out.
At every reference optimisation, the reference weight is calculated anew. As the addi-
tional pieces increase the base for the calculation, the reference also becomes more
exact.
22
All manuals and user guides at all-guides.com
0
1. 8 33
0
1. 8 33
Reference weight
] display will appear in the reference weight
Placed number of pieces insufficient for reference calculation
CXB models < 40 d
CXP models < 20 d
Placed reference weight insufficient for reference calculation
CXB models < 4/5 d
CXP models < 1/5 d
] display disappears.
0
M+
0
M+
CXB/CXP-BA-e-0812

Publicité

loading