Other Functions

7.1
Setting the Auto Power-Off Timer
1.
Switch on the scales.
2.
Hold both
and
3.
Set the timer using the numeric keys (timer range from 0 ~ 30 min). The scales will not switch off
automatically when the timer is set to 0.
4.
Press
to confirm your choice.
Note: This function is only valid when powering the scales with a battery.
7.2
Calibration
1.
Switch on the scales.
2.
Hold both
and
3.
Calibrate the zero point first. The scales display unit.0 in the PIECE WEIGHT window. Make sure there
is no object on the platform. Press
4.
The scales display unload in the PIECE WEIGHT window. Press
5.
The scales will now display load in the PIECE WEIGHT window. Place a reference weight onto the
platform. The reference weight must be 25% ~ 100% FS. Press
stabilised.
6.
The scales display inp.ld in the PIECE WEIGHT window. Now, enter the reference weight's value using
the numeric keys and press
7.
Remove the reference weight from the platform and press
8.
The scales will now automatically reset and return to the normal weighing mode. Exit the calibration
mode by pressing
8.

Display Description

- - - - - -
Overload or overrange
Uncertain value
- - - - - -
- - - - - -
Weight signal is too small
Zero point is too high
0- - - - -
Zero point is too low
0- - - - -
EEPROM cannot be accessed
Eep.er0
The EEPROM parameters differ from the backup data
Eep.er1
The parameter has not been set
Eep.er2
Calibration error
Cal.err
Input signal over maximum range
Ad- - - -
Input signal under minimum range
Ad- - - -
Capacity-related data display
Cap
Voltage-related data display
Uol
Totalised data
Add
Piece weighing
Pc.t
Sample quantity
Spl.pcs
Upper limit
Up.pcs
Lower limit
Lo.pcs
V. 02 – 07/07/2015
keys pressed to enter the setting mode.
keys pressed to enter the calibration mode.
.
to confirm.
. A calibration error will be reported by cal.err.
VTBAL6
to calibrate the zero point.
to confirm the zero point again.
5
when the displayed value has
©Velleman nv