10. Declaration of Electromagnetic Compatibility (EMC)
Essential Performance
When operated at maximum suction (Level 8, Mode 3) the vacuum pressure must not
exceed its specification.
EN 60601-1-2 STANDARD TEST REQUIREMENTS
Immunity Test Levels for ENCLOSURE PORT
PHENOMENON
Electrostatic
Discharge
Radiated RF EM
fields
a)
Proximity fields
from RF wireless
communications
equipment
RATED power
frequency magnetic
fields
d) e)
Immunity Test Levels for Input AC Power PORT
PHENOMENON
Electrical fast
transients/bursts
a) l) o)
Surges
a) b) j) o)
Line-to-line
Surges
a) b) j) k) o)
Line-to-ground
Conducted
disturbances
induced by RF fields
c) d) o)
Voltage dips
f) p) r)
Voltage
Interruptions
f) i) o) r)
BASIC EMC
STANDARD OR
TEST METHOD
IEC 61000-4-2
IEC 61000-4-3
IEC 61000-4-3
IEC 61000-4-8
BASIC EMC
STANDARD OR
Professional Healthcare
TEST METHOD
IEC 61000-4-4
IEC 61000-4-5
IEC 61000-4-5
IEC 61000-4-6
IEC 61000-4-11
IEC 61000-4-11
IMMUNITY TEST LEVELS
Professional Healthcare
Facility Environment
± 8 kV contact
± 2 kV, ± 4 kV, ± 8 kV, ± 15 kV air
3 V/m
f)
80 MHz – 2,7 GHz
b)
80 % AM at 1 kHz
c)
50 Hz or 60 Hz
IMMUNITY TEST LEVELS
Facility Environment
100 kHz repetition frequency
± 0,5 kV, ± 1 kV
± 0,5 kV, ± 1 kV, ± 2 kV
3 V
m)
0, 15 MHz – 80 MHz
6 V
in ISM bands
m)
between 0,15 MHz and
80 MHz
n)
80 % AM at 1 kHz
e)
0 % UT ; 0,5 cycle
At 0°, 45°, 90°, 135°, 180°, 225°, 270° and 315°
0 % UT ; 1 cycle and
70 % UT ; 25/30 cycles
Single phase: at 0°
0 % UT ; 250/300 cycles
Home Healthcare
Environment
10 V/m
80 MHz – 2,7 GHz
80 % AM at 1 kHz
See 8.10
30 A/m
g)
Home Healthcare
Environment
± 2 kV
3 V
m)
0, 15 MHz – 80 MHz
6 V
in ISM and amatuer
m)
radio bands between 0,15
MHz and 80 MHz
80 % AM at 1 kHz
g)
h)
h)
f)
b)
c)
n)
e)
q)
17