Télécharger Imprimer la page

Nilfisk VHS110 Manuel D'instructions page 31

Publicité

Reinstall the deck and lock it with the closing band (1).
Upstream absolute filter replacement
Figure 10
Clean the primary filter as described in the "Filter cleaning"
section.
Release the closing band (1).
Lift the deck and disconnect the grounding braid (2) via
the knob (3).
Remove the absolute filter (4) by unscrewing the nut (5).
Cover the absolute filter (4) with a plastic bag, close the
bag tightly and dispose of the filter in accordance with
current legislation.
Insert a new absolute filter (4) with the same filtering
characteristics as the removed one.
Lock the absolute filter with the nut (5).
Reconnect the grounding braid (2) via the knob (3).
Reinstall the deck and lock it with the closing band (1).
Replacing the motor cooling filters
Figure 11
Adjust the tightening devices and remove the covers (1)
Remove the absolute filter (2) and the related gasket (3).
Cover the absolute filter (2) with a plastic bag, close the
bag tightly and dispose of the filter in accordance with
current legislation.
Insert a new absolute filter (2) and the related gasket (3)
the same filtering characteristics as the removed one.
Refit the covers (1) and lock them into place, using the
tightening devices.
Autoclavable parts
This vacuum cleaner can be constructed with autoclavable
parts. All stainless steel parts and silicone rubber gaskets can
be cleaned / decontaminated by means of an autoclave at 121
C – this will secure an absolute decontamination of the parts.
The autoclavable parts are:
Trolley (without wheels)
Dust container (with wheels and release handle)
Filter chamber
Screws, nuts etc.
Silicone rubber gaskets
Inlet (without seal to be replaced)
All accessories: Hoses, nozzles and white silicone hose.
All NON autoclavalbe parts shall be cleaned by means an
appropriate detergents / sanitizers, to ensure an absolute clean
surface and to prevent any bacterial growth or the like.
The NON autoclavable parts are:
Motor unit
Machine wheels fixed to the trolley
Filter chamber with PullClean system
Knobs for the filter chamber release
Plastic parts
Filter
Bag
Tightness inspection
Hoses check
Make sure that connecting hoses are in a good condition and
correctly fixed.
If the hoses are damaged, broken or badly connected to the
unions, they must be replaced.
When sticky materials are treated, check for possible clogging
along the hose, in the inlet and on the baffle plate inside the
filtering chamber.
To clean, scrape the inlet from the outside to remove deposits.
Filtering chamber gasket check for machines
equipped with dust container
If the gasket (1) between the container and the filtering chamber
(3) fails to guarantee sealing:
Loosen the four screws (2) that lock the filtering chamber
(3) against the machine structure.
Allow the filtering chamber (3) to lower down and tighten
the screws (2) once it has reached the sealing position.
If an optimal seal is not yet obtained or if there are tears, cracks,
etc., the gasket must be replaced.
Filtering chamber gasket check for machines
equipped with Longopac® system
Ensure that the Longopac® bag is tight with the gasket (2).
Also check the seal of the gasket positioned on the discharge
clapet (1).
The gasket must be replaced if it is torn, cut, etc...
13
VHS110
Figure 12
Figure 13
IT
GB
F
D
E
C423

Publicité

loading