Analyzer and Bench Meter
1.3. Touch screen
1
9
10
11
1.
Date and time
2.
Clock icon (a-AB41PH): appears when the timing function is used.
Note: this function is only available for a-AB41PH model.
3.
Computer icon: appears when connects to a computer.
4.
Auto endpoint icon
5.
Measurement icon: blinks when measurement is in progress.
6.
Pinter icon: touch to print.
7.
Calibration icon: touch to calibrate; blinks when calibration is in progress.
8.
Menu icon: touch to enter menu settings.
9.
PH electrode condition (a-AB33M1, a-AB33PH and a-AB41PH)
Slope: 95-105%
and offset: ± (0-15) mV.
Electrode condition is
good.
2
3
4
16
Figure 1-1 Touch screen
; Continuous reading icon
5
6
17
Slope: 90-95%
or offset: ± (15-35) mV.
Electrode condition is
acceptable.
7
8
18
.
Slope: less than 90%
or offset: ± (35-60) mV.
Electrode condition is not
good or needs cleaning.
EN-5