External Voltage Inputs (continued)
Set voltage source to 10 V. Note: the error message SHrt indicates a bad connection.
U=10
Off
Sure
External Current Inputs
Display
I+
Switch on Power, Calibrate link In CAL Position.
Off
u=0
Connect current source and Set to 0 mA.
On
I=0
Set current source to 20 mA. Note: the error message Open indicates a bad connection.
I=20
On
Sure
8
Ó
Off
On
Ó
I-
V
Keys
Off
On
Ô Õ Ù Ó
Off
Off
Ó
Ó
Off
Off
Ó
Calibrate 10 V point
To Save Values (or any other key to restart).
Action
Press sequence to select External Voltage
mode. INC and DEC Leds will both light.
Calibrate 0 mA point
Calibrate 20 mA point
To Save Values (or any other key to restart).