Weidmüller Portacal 1000 Mode D'emploi page 12

Appareil d'étalonnage pour signaux de courant/tension
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Ò
switches operating modes. If you are adjusting a value you will
lose those changes.
Õ
toggles between increment and decrement mode.
Ù
â
á
,
and
Calibrate/Run mode selection jumper
Place the calibration jumper in the CAL position to start calibrating. You
will have to open the case to select the jumper and normal precautions
against ESD are essential. Take careful note of the position of the wires
connecting the front and back of the case together.
Voltage Source Mode
Display
I+
Switch on Power, Calibrate link In CAL Position, Measure output voltage
u=0
u=0
Off
u=10
u=10
Sure
Current Source Mode
Display
I+
Switch on Power, Calibrate link In CAL Position
Off
u=0
Connect multimeter to measure output current (mA).
6
adjust the output,
I-
V
Keys
Õ Ù â á
Ó
Off
On
Õ Ù â á
Ó
Ó
I-
V
Keys
Off
On
Ò
á
provides the finest adjustment.
Action
Adjust Output Voltage to 0 V
Accept output
Adjust Output Voltage to 10 V
Accept output
To Save Values (or any other key to restart).