5) Calibrate the probes with pH and Redox calibration solution
A timer will stop after 120 seconds and notify you if the calibration was successful.
Start the calibration of the Redox probe by pressing the Redox calibration key (bottom key) for
10 sec and perform the above operation, now for Redox.
5) Insert the probes into the dosing tube, disconnect the installation from the mains,
connect water and test for leaks.
6) Set the parameters:
The settings are made by the buttons on the blue lid
Middle button is the enter button, top and bottom buttons are scroll and selection buttons
1 LANGUAGE :
You choose the desired Language
2 DOSING METHOD:
Attention!
With salt electrolysis you may NOT use the dosing methods Continuous and proportional.
For salt electrolysis you can choose 2 options:
Option 1: Salt time or
Option 2: Salt Rx, choice goes to Salt RX for Start-up.
Choose option 2 : Salt RX.
3) PH+ or PH-
PH- (min) must always be set when using the dosing unit on salt electrolysis
It is not possible to add both PH+ and PH-: PH- or PH+, never both at the same time.
4) PH SP(SetPoint)
As a rule, the pH will be regulated at 7.4
5) Rx SP(Redox SetPoint)
At startup, we recommend setting the setpoint to 700mV. When this set point is reached,
measure the chlorine using a reliable staining method. If the chlorine content is too low,
increase the set point. If the chlorine content is too high, lower the setpoint. ATTENTION: with
salt electrolysis it sometimes happens that the RX measurement (temporarily) does not work.
Read the manual for more details. Regularly measure the chlorine level during the start-up
phase
6) Flow switch yes, Level switch no or yes
All Salt electrolysis units come with a flow switch. Units ZLAX0017-P and 0018-P come with a
level switch which will stop dosing pH liquid if the container is empty.
7) Safety alarm: on or off or info
Immerse the probes in the pH (pH 7) and Redox
(465-468 mV) calibration fluids and wait 5
minutes to equilibrate before continuing with the
calibration process.
Press the pH calibration button (top button) for
10 seconds and then select Calibration via buffer
pH7