TEMP
• Disconnect test lead probes from voltage source and meter.
• Move Input Jack Locks to "TEMP" setting.
• Use K-Type thermocouple temperature probes only.
• Stated accuracy does not account for thermocouple accuracy.
Features:
Range
-328° to 999°F
(-200° to 999°C)
1000° to 2462°F
(1000°C to 1350°C)
The new DL579 clamp meter offers a digital field temperature calibration
procedure to make the process easier. Here are the steps.
1) Set dial position to "°F°C"
2) Press and hold SELECT button for about 2 seconds
3) Press HOLD button while "FCAL" is displayed on top LCD to enter
Field Calibration.
4) After immersing temperature probe in ice bath and temperature is stable,
press and hold Hz/Duty button
Temperature °F/°C
RANGE
MAX/MIN
AUTO
AUTO
°F
True RMS
CLAMP METER
SELECT
INRUSH
DISCONNECT TEST LEADS
TEMP
CATIV 300V
CATIII 600V
SELECT
BT
Resolution
0.1°F
(0.1°C)
1°F
(1°C)
Temperature Calibration
Temp
• Press SELECT = ˚F or ˚C
• Readings show on lower display
Accuracy
±(1.0% + 3.6°F)
±(1.0% + 2.0°C)
±(1.0% + 3°F)
±(1.0% + 2°C)
11
Overload Protection
30V RMS