• The batteries shall be checked before use and be
replaced if necessary.
• Hold the voltage tester and accessories by the
designated probe handle areas only, the LCD screen
and LED indicator must not be covered. Do not touch
the test probes in any case before and during test.
• The voltage tester may be used only within the
specified measurement ranges and in low-voltage
installations up to 690V AC/DC (2100-Gamma: up to
1000 VAC / 1200 VDC).
• The voltage tester may be used only in the
overvoltage category for which it has been designed
for!
• Before and after use, always check that the voltage
tester is in perfect working condition (e.g. check on a
known voltage source or on a proving unit).
• The functioning of the voltage tester shall be checked
shortly before and after a test. If indication of one or
more steps fails, or no functioning is indicated, the
voltage tester must be taken out of operation.
• The voltage tester complies with protection degree
IP 64 (splash-proof and dust-tight) and therefore can
also be used under humid conditions.
• The voltage tester operates correctly only within a
temperature range of -15°C to +55°C at relative air
humidity less than 95% (without condensation).
• If the safety of the user cannot be guaranteed, the
voltage tester must be taken out of operation and
secured against unintentional use.
• Safety is no longer guaranteed in the following cases:
- obvious damage
- if the voltage tester can no longer perform the
required measurements/tests
- stored for too long in unfavorable conditions
- damaged during transport
- leaking batteries
• For all work, observe the accident prevention
regulations of the professional trade association for
electrical installations and equipment and/or other
local safety regulations.
• Unauthorized persons shall not disassemble or
assemble the voltage tester and supplementary
equipment. The voltage tester may be serviced by an
Amprobe authorized service technician only.
• Operational safety is no longer guaranteed if the
voltage tester is modified or altered.
Before Using the Tester
W
Before conducting any test, follow the safety
instructions. Before using the voltage tester, always
perform a function test.
W
Disconnect the voltage tester completely from any
measuring circuit first.
The "GS 38 protective cap"
before the tests. To do this, pull them forward off the test
probes.
W
The "GS 38 protective cap"
national regulations or directives.
Switching ON / OFF the Tester / Auto Power OFF:
By activating the push button
be switched ON without applying any voltage to the test
probes. The torch light is now ON and the tester is also
ready to test continuity (for 2100-Beta and 2100-Gamma,
the LCD screen is also ON and shows "---").
If no voltage is applied to the test probes and no
continuity is detected, the voltage tester and torch light
automatically switches OFF after approx. 30 seconds.
The voltage tester can be switched OFF manually by
pressing the push button
can be removed
15
may be required by
13
12
, the voltage tester can
for 5 seconds.
12
6