Chapter II
3.2.4. Use of 2 power supplies in series with the "TRACKING" function (symmetrical
at 2.5 A max. (AX 502, AX 503 and AX 503F)
 Press in the "TRACKING" button ("ON", LED lit).
The pushbutton is set to "ON" sets up a series connection between the "-" terminal of the
"MASTER" power supply and the "+" terminal of the "SLAVE" power supply.
Procedure:
 In sequence (power supply a and then b), adjust the "current " potentiometers to define
identical limit currents.
 Connect up to the circuit to be powered (external terminals of the "MASTER" and "SLAVE"
power supplies).
 The "MASTER" source is on the left (a); set the required initial voltage.
 The "SLAVE" source is on the right (b).
 In this function, only the voltage controls of the "MASTER" source (a) will be used from now on.
 The symmetrical "SLAVE" source (b) reacts to the "MASTER" source (a) voltage variations by
proportional "tracking" variations in absolute values ("TRACKING" function).
For the use of symmetrical voltages (e.g. ± 15 V), one of the 2 terminals (MASTER "-"
or SLAVE "+") will be connected to the circuit to be powered ("0" point).

V "MASTER" (a)

V "SLAVE" (b)
 
Example
If the "SLAVE" source (b) is taken outside its limits (voltage or current) by the "MASTER"
source (a), the corresponding "LIMIT" LED lights up.
3.2.5. Use of the 2.7 V - 5.5 V (5 A) power supply (AX 503 only)
 If necessary, connect the "
 Connect the "+" and "-" terminals to the points of the circuit to be powered using the connection
leads (*).
 Set the required voltage between 2.7 V and 5.5 V using the adjustment potentiometer.
 For "fine" adjustment, connect these terminals to a multimeter (appropriate calibre).
 If the output current reaches the 5 A limit current value, the "LIMIT 5 A" LED comes on, the
power supply automatically switches to "current" regulation mode and the voltage is not longer
regulated.
Application:
(*) AX 503F
Stabilized power supplies
Action:  V "MASTER" (a)
Reaction:  V "SLAVE" (b) (absolute value)
V "MASTER"
V "SLAVE"
V "MASTER"
V "SLAVE"
V "MASTER"
V "SLAVE"
" earth terminal to the circuit to be powered (*).
This output is designed to power logic circuits at 3 or 5 volts with consumption
up to 5 A.
AX 503F : power supply of electronic circuits, leds, Zener diodes ...
Initial
V "MASTER" /
conditions
V "SLAVE"
+12 V
(a)
1 / 1
-12 V
(b)
+24 V
(a)
2 / 1
-12 V
(b)
+12 V
(a)
1 / 2
-24 V
(b)
30 V sources)
 V
causes
"MASTER"
+3 V
+3 V
-3 V
-4 V
-4 V
+2 V
+2 V
+2 V
-4 V
Results
+15 V
-15 V
+20 V
-10 V
+14 V
-28 V
25