5.2.
DC VOLTAGE GENERATION
The maximum DC which can be applied to inputs is 30V DC. Do not
measure voltages exceeding the limits given in this manual. Exceeding
these limits could result in electrical shocks to the user and damage to the
instrument.
1. Press the MODE key and select modes OUT V or OUT mV. The symbol "OUT" is
shown on the display.
2. Use the adjuster knob (see § 4.2.6), the 0-100% key (see § 4.2.2) or the 25%/
(see § 4.2.3) to set the desired value of output voltage. The maximum values available
are 100mV (OUT mV) and 10V (OUT V). The display shows the value of voltage
3. Insert the green cable into input lead mV/V and the black cable into input lead COM.
4. Position the green lead and the black lead respectively in the points with positive and
negative potential of the external device (see Fig. 4)
5. To generate a negative voltage value, turn the measuring leads in the opposite
direction with respect to the connection in Fig. 4
CAUTION
Fig. 4: DC voltage generation
EN - 9