User manual of Colson pulse Oximeter OxyPad – Home ref. CMS50DL / Pro ref. CMS50D
Guidance and manufacture's declaration – electromagnetic immunity –
for EQUIPMENT and SYSTEMS that are not LIFE-SUPPORTING
Guidance and manufacture's declaration – electromagnetic immunity
The oximeter is intended for use in the electromagnetic environment specified below. The customer or the user of the oximeter should
Immunity test
IEC 60601 test level
Radiated RF
IEC 61000-4-3
NOTE 1
At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2
These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection
from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios,
amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the
electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured
field strength in the location in which the oximeter is used exceeds the applicable RF compliance level above, the oximeter
should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary,
such as reorienting or relocating the oximeter.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
assure that it is used in such an environment.
3 V/m
80 MHz to 2.5 GHz
Compliance
Electromagnetic environment - guidance
level
Portable and mobile RF communications equipment should
be used no closer to any part of the oximeter, including
cables, than the recommended separation distance
calculated from the equation applicable to the frequency of
3 V/m
Where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended separation
Field strengths from fixed RF transmitters, as determined
by an electromagnetic site survey,
Interference may occur in the vicinity of equipment
17
the transmitter.
Recommended separation distance
3
5 .
=
d
V
1
3
5 .
=
d
P
80 MHz to 800 MHz
E
1
7
=
d
P
800 MHz to 2.5 GHz
E
1
distance in metres (m).
compliance level in each frequency range.
marked with the following symbol:
P
a
should be less than the
b