User manual of Colson pulse Oximeter OxyPad – Home ref. CMS50DL / Pro ref. CMS50D
9 Troubleshooting
Trouble
The SpO
and
2
Pulse Rate cannot
be displayed
normally
The SpO
and
2
Pulse Rate are not
displayed stably
The device can
not be turned on
The display is off
suddenly
Possible Reason
1. The finger is not properly
positioned.
2. The patient's SpO
be detected.
1. The finger is not placed inside
deep enough.
2. The finger is shaking or the
patient is moving.
1. The batteries are drained or
almost drained.
2. The batteries are not inserted
properly.
3. The malfunction of the device.
1. The device will power off
automatically when it gets no
signal within 5 seconds.
2. The batteries are almost
drained.
1. Place the finger properly and
try again.
is too low to
2. Try again; Go to a hospital for
2
a diagnosis if you are sure the
device works all right.
1. Place the finger properly and
try again.
2. Let the patient keep calm
1. Change batteries.
2. Reinstall batteries.
3. Please contact the local
service center.
1. Normal.
2. Change batteries.
12
Solution