dual ruby light, infrared heater, direct sunlight and etc.
6.
Strenuous action of the subject or extreme electrosurgical interference may also affect
the accuracy.
7.
Testee cannot use enamel or other makeup.
3.3 Clinical restrictions
1.
As the measure is taken on the basis of arteriole pulse, substantial pulsating blood flow
of subject is required. For a subject with weak pulse due to shock, low ambient/body
temperature, major bleeding, or use of vascular contracting drug, the SpO
(PLETH) will decrease. In this case, the measurement will be more sensitive to
interference.
2.
For those with a substantial amount of staining dilution drug (such as methylene blue,
indigo green and acid indigo blue), or carbon monoxide hemoglobin (COHb), or
methionine (Me+Hb) or thiosalicylic hemoglobin, and some with icterus problem, the
SpO
determination by this monitor may be inaccurate.
2
3.
The drugs like dopamine, procaine, prilocaine, lidocaine and butacaine may also be a
major factor blamed for serious error of SpO
4.
As the SpO
value serves as a reference value for judgement of anemic anoxia and
2
toxic anoxia, some patients with serious anemia may also report good SpO
measurement.
measure.
2
waveform
2
2