4.2. Control commands
4.2.46 QAD
Request to send the information rate down 'analogue on/off' via RS232 interface.
For example:
4.2.47 QDS
Request to send the dispensing volume via RS232 interface.
For example: "1.275 'CR' 'LF'"
Transmission is only possible in modes DIS R and DIS C.
In other modes, string "not defined
4.2.48 QPI
Request to send the pipetting volume via RS232 interface.
For example: "1.275 'CR' 'LF'"
Transmission is only possible in modes PIP and DIL.
In other modes, string "not defined
4.2.49 QLI
Request to send the limit volume via RS232 interface.
For example: "1.275 'CR' 'LF'"
If V-LIM is 'off', the string "OFF 'CR' 'LF'" is transmitted.
Transmission is only possible in modes DOS and DIS C.
In other modes, string "not defined
4.2.50 QDL
Request to send the diluting volume via RS232 interface.
For example: "1.275 'CR' 'LF'"
Transmission is only possible in mode DIL.
In other modes, string "not defined
4.2.51 QUN
Request to send the unit via RS232 interface.
For example: "mg/l 'CR' 'LF'"
Transmission is only possible in mode DOS.
In other modes, string "not defined
4.2.52 QAF
Request to send the information 'auto fill on/off' via RS232 interface.
For example:
44
Query rate down 'analogue on/off', live
Analogue on
"on 'CR' 'LF'"
Analogue off " off 'CR' 'LF'"
Query dispensing volume, live
'CR' 'LF'" is transmitted.
Query pipetting volume, live
'CR' 'LF'" is transmitted.
Query limit volume, live
'CR' 'LF'" is transmitted.
Query diluting volume, live
'CR' 'LF'" is transmitted.
Query unit, live
'CR' 'LF'" is transmitted.
Query 'auto fill on/off', live
auto fill on
"on 'CR' 'LF'"
auto fill off "off 'CR' 'LF'"
765 Dosimat