Spark (Universal Mount)
Spark Standard
Spark TruColor
Light Output
32 Lumens
32 Lumens
Color Rendering Index (CRI)
CRI 70+
CRI 90+
Correlated Color Temperature (CCT)
CCT 6500K
CCT 5700K
Weight of headlight with filter
.56 ounces (15.8 grams)
Diameter of headlight
.6 inches / 14 mm
Weight of battery pack
.63 ounces (17.9 grams)
Dimensions of battery pack
Length = 1.7 inches (43 mm) | Width = 1.1 inches (28 mm) | Height = 0.5 inches (13 mm)
Charge time (each battery pack)
3 Hours
Run time (each battery pack)
3 Hours
Battery type
Lithium-ion / 3.7v / 425 mAh
Electrical
Class II
Operation mode
Continuous
Standards
EMC = IEC60601-1-2, class B, EN55011 | Safety = IEC60601-1, IEC 62471
Operating environment
32° to 77°F (0° to 25°C) | 10% to 85% relative humidity (non-condensing) | 0.5 atm to 1.0 atm (500 hPa to 1060 hPa) atmospheric pressure
Problem
Headlight does not turn on.
Battery indicator light is blinking scarlet.
Battery indicator light does not illuminate amber while charging.
14
INSTRUCTIONS FOR USE AND CARE
Orascoptic
Specifications
Spark (Through-the-lens Mount)
Spark TruColor (Generation 2)
Spark Standard
Spark TruColor
40 Lumens
32 Lumens
32 Lumens
CRI 92+
CRI 70+
CRI 90+
CCT 6500K
CCT 6500K
CCT 5700K
.43 ounces (12.2 grams)
Troubleshooting
Solution
Check that the battery pack is charged. Remove and reattach the battery pack, making sure both locking tabs are secure. Attach the other battery pack to determine if the
problem is in the headlight or battery pack.
Indicates the battery is low. Charge the battery.
Check that the Micro-USB plug is secured in the charging port on the back of the battery pack. Check that the USB plug is secured in the power adapter. Check that the electrical outlet is
producing power.
™
™
Spark
Series Headlight
CISPR 11 - Emissions: Radiated RF
Spark TruColor (Generation 2)
IEC 61000-4-2 - Immunity: Electrostatic Discharge
40 Lumens
CRI 92+
IEC 61000-4-3 - Immunity: Radiated RF
CCT 6500K
IEC 61000-4-3 - Immunity: Proximity Fields from RF Wireless
Communications Equipment
IEC 61000-4-8 - Immunity: Power Frequency Magnetic Fields
Electromagnetic Compatibility Test
Group 1, Class B
±8kV contact
±15kV air
3V/m, 80MHz – 2.7GHz, 80% 1kHz AM
IEC 60601-1-2:2014, section 8.10
30A/m, 50/60Hz