Performing tests in manual or production line modes
1.
Turn on the device with the On / Off key (
2.
The display will show the AUTO VERIFICATION message and then PRESS START.
3.
Select the range and current (
4.
Press the
key.
5.
The message LOW CURRENT will appear. Turning the control
the current until reaching the desired stable value measured on the display or in the bargraph
current indicator.
6.
When the current value is greater than 50% of the nominal value on the 1 mA range and 10%
on the other ranges, the display will indicate the measured resistance value.
7.
Upon completion of this process the display will indicate the measured resistance value and
other test information.
8.
This information can be saved in the machine's internal memory during and / or after the test
simply by pressing the
9.
To finish the test, press the
without pressing the
10. Finally, after all measurements have been completed, switch the device off using the on / off
switch (
).
Never connect or disconnect the test leads with the equipment performing tests. If you
have to make any changes to the connection, it must be done with the equipment in
stand-by (not generating current) or turned off.
key.
key. Do not turn off the product with the power switch (
key.
14
).
).
clockwise will increase
)