Front panel
2.3.
pressure Adjustment Knob
(1)
The pressure adjustment knob controls the air pressure output. When turning
clockwise, the output pressure will be increased. On the other hand, when turning
counter-clockwise, the output air pressure will be decreased.
NOTE: You can also check if the pressure is suitable for the patient
by sliding one hand beneath the air cells at the level of the patient's
buttocks. The air cells will alternately inflate and deflate.
should feel slight contact with the buttocks when the air cells be-
neath the buttocks deflate.
Low Pressure Indicator
(2)
When the mattress is initializing, the Low Pressure LED will illuminate until the
appropriate pressure is reached (according to the Pressure Adjustment Knob); this is
normal. Otherwise, the Low Pressure LED is a warning, indicating that the pressure
in mattress is unusually low. Check to ensure all connections are secured and
correctly installed as per the installation instructions.
Normal Pressure Indicator
(3)
When the green LED illuminates, the pressure inside of air mattress has reached the
desired pressure setting (preset on the Pressure Adjustment Knob).
Main Power Switch
(4)
To turn on/off the pump unit with illuminated green light.