 The extension cable accessory (P/N: 30429931) cannot be used if the scale will be used in a legal for
trade or legally controlled application, because there is no provision to seal the connection between the
indicator and platform.
Type1. DxxxRxx Bases
DxxxWxx Bases
Selecting the Location
Level the base by adjusting the four leveling feet until the bubble in the level indicator is centered.
Note: Ensure that the base is level each time its location is changed.
MAINTENANCE
Cleaning
The base components should be kept clean and free of excessive material build up.
A damp cloth with water and a mild detergent may be used to wipe clean the external surfaces – do not
use acids, alkalis, strong solvents or abrasive materials.
To maintain surface integrity, do not allow debris to remain on the surface of the painted or stainless steel base
frame.
Remove the pan and use fresh water to clean the base frame surface of any debris. Dry the base with a
clean cloth or allow to air dry before replacing the pan. Repeat daily for best care.
Troubleshooting
An Ohaus product service specialist will be available to provide assistance. Outside the USA, please visit our web
site to locate the Ohaus office nearest you.
TECHNICAL DATA
Technical Specifications
The technical data is valid under the following ambient conditions:
Ambient temperature: -10ºC to +40 ºC
Relative humidity: 10% to 90% relative humidity, non-condensing
Altitude: Up to 2,000 m
Operability is assured at ambient temperature -10ºC to +40 ºC
Note: Not all models may be available in your area. Check with your local Ohaus representative.
TABLE 1. LOAD CELL CONNECTION
FUNCTION
WIRE COLOR
+ Excitation
Green
- Excitation
Black
+ Signal
White
- Signal
Red
+ Sense
Blue
- Sense
Brown
Shield
Yellow
CORRECT
INCORRECT