Floorline LTC
Guidance and manufacturer´s declaration
Recommended Separation Distances between portable and mobile RF communications equipment and the Medical Bed
The Medical Bed is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The
customer or the user of the Medical Bed can help prevent electromagnetic interference by maintaining a minimum distance
between portable and mobile RF communications equipment (transmitters) and the Medical Bed as recommended below,
according to the maximum output power of the communications equipment.
Rated
Separation Distance
Maximum
according to frequency of transmitter
Output Power
( m )
of transmitter
150 kHz to 80 MHz
(W)
0.01
0.12
0.1
0.37
1
1.17
10
3.69
100
11.67
For transmitters rated at a maximum output power not listed above, the recommended separation distance (d) in metres (m) can
be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of
the transmitter in watts (W), according to the transmitter manufacturer.
NOTE 1
At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2
These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from
structures, objects and people.