7.
Reinsert plug/cord into wall socket.
8. Turn ON the wall switch (in countries
where this is applicable) and leave
permanently on, for charging back‐up
battery.
Mattress Platform Positions
• True Floor-Level - 95mm (3¾") off floor
• Highest Platform Height 700mm (27½")
• Backrest Panel - 70°
• Thigh Panel - 45° }operate together as
• Calf Panel - 25° }the kneebreak panel
• Kneebreak Position - 110°
• Trendelenburg Tilting Position - 18°
• Reverse-Trendelenburg Tilt Position - 18°
• Head High Trendelenburg Tilt Position - 45°
• Head Low Trendelenburg Tilt Position - 45°
The Handset
1. Back rest down
2. Hi-lo down
3. Knee break down
4. Auto contour down
5. Trendelenburg tilt
6. Back rest up
7. Hi-lo up
8. Knee break up
9. Auto contour up
10. Reverse Trendelenburg tilt
The Handset is directly connected to the control
box, via a curly cord. Each button has a Raise
(Left Side) and a Lower (Right Side) function
that adjusts all mattress platform positions.
NOTE! All buttons should be explained to
patients.
6
1
3
9
4
7
2
10
5
Lockout
A small blue plastic 'lock out key' is included, to
lock any function on the bed:
11. Insert the key into the two holes in the
button you wish to lock out
12. Gently turn the key from the 'green' - 12
o'clock position toward the unlock or
lock symbols at 11 or 1 o'clock positions
respectively.
Gently turn the lockout key in the holes, or key
tabs may break.
To immediately stop movement release
the handset button!
• Only press one button at a time. When
desired position is reached, release the
button.
• When maximum raised or lowered positions
8
of mattress base or backrest are reached,
a built-in limit switch will automatically
override the handset button and movement
will stop.
Do not continue to press any
button after desired position is
attained!
• Movement in the opposite direction will be
resumed when the appropriate button is
pressed.
• The handset may be stored by clipping it
into the mattress retainers (side of mattress
platform) or over the Head/Foot Ends.
• No Lockout function or key is available with
this handset type.
www.humancaregroup.com