TECHNICAL SPECIFICATIONS OF DEVICE
Model:
P0711EM-1 F400 combined RF9-1
Operating pressure (with neb.):
RESPIRATORY
SYSTEM
CONFIGURATION
BOWL
MMAD (μm)
Output rate (ml/min')
Respirable fraction < 5
μm (FPF)
(1) In vitro characterisation performed by TÜV Rheinland Italia S.r.l. in cooperation with the University of Parma. More
details are available on request.
(2) Data collected according to Flaem's internal procedure.
(3) Data calculated on the basis of the values obtained with Malvern Spraytec, and compared with the values of the
tests performed at the University of Parma.
APPLIED PARTS
Type BF applied parts are:
Weight:
ENVIRONMENTAL CONDITIONS
Operating conditions:
Ambient temperature
Relative air humidity
Atmospheric pressure
Storage and transport conditions:
Ambient temperature
Relative air humidity
Atmospheric pressure
0.65 bar.
UPPER
AIRWAYS
Oropharyngeal
Tracheobronchial
tract
Without
Selector in
Selector
position 1
7.71
5.12
(3)
0.53
0.36
(2)
26.7%
48.8%
(3)
patient accessories (B3, B4, B5)
1.300 Kg
Between +10°C and +40°C
Between 10% and 95%
Between 69 KPa and 106 KPa
Between -25°C and +70°C
Between 10% and 95%
Between 69 KPa and 106 KPa
8
LOWER AIRWAYS
Pulmonary
tract
tract
Selector in
position 2
B2.1c
3.67
(1)
(1)
0.30
(2)
(2)
63.4%
(1)
Peripheral
tract
Selector in
position 3
B2.1c
2.49
(1)
0.23
(2)
83.1%
(1)
(1)
E
Th
A
B
C