2. Apply FORA control solution
Shake FORA control solution vial thoroughly before use. Squeeze out
a drop and wipe it off, then squeeze out another drop and place it on
the tip of the vial cap. Hold the meter in an angled position so that the
absorbent hole of the test strip touches the drop. Once the confirmation
window fills completely, the meter will begin counting down. To avoid
contaminating the control solution, do not directly apply FORA control
solution onto the strip.
3. Read and compare the results
After the meter counts to 0, the result of the control solution test will
appear. Compare the result with the range printed on the test strip vial.
The result should fall within this range.
Out-of-range results
If you continue to have test results fall outside the range printed on the
test strip vial, the meter and strips may not be working properly. Do NOT
test your blood. Contact customer service or place of purchase for help.
NOTE:
• The control solution range printed on the test strip vial is for FORA control
solution use only. It is not a recommended range for your blood glucose level.
• See the Meter Maintenance section for important information about your FORA
control solutions.