6.8 Linearization (non-verified models only)
Linearity shows the greatest deviation of a weight display on the scale to the value of
the respective test weight according to plus and minus over the entire weighing
range.
If linearity deviation is discovered during a testing instrument control, you can
improve this by means of linearization.
•
Carrying out linearization is restricted to specialist staff possessing well
acquainted with the workings of weighing scales.
•
The test weights to be used must be adapted to the weighing scale's
specifications; see chapter 3.4 "testing instruments control".
•
Observe stable environmental conditions. Stabilisation requires a
certain warm-up time.
•
After successful linearization you will have to carry out calibration; see
chapter 3.4 "Testing instruments control"
Tab. 1: Adjustment points
Adjustment
weight
1.
2.
3.
4.
Operation
How to carry out linearization:
Switch on balance and during self-check
press the TARE key
Use the numeric keys to enter password
„9999"
Confirm with TARE button
CPB-N / CPB-DM-BA-e-1625
CPB 6K0.1N
CPB 15K0.2N
1 kg
2 kg
4 kg
6 kg
CPB 30K0.5N
2.5 kg
5 kg
10 kg
15 kg
Display
5 kg
10 kg
15 kg
30 kg
19