11. TECHNICAL DATA
Operating voltage .................................100 - 240 V/AC, 50/60 Hz
Current consumption ............................Max. 1.1 A
Protection class ....................................II
DC-output .............................................1
Connection sockets for DC-output ........4 mm round sockets
Adjustable output voltage .................... Area 1: 0 - 16 V/DC
Adjustable output current ..................... Area 1: 0 - 5 A
Resolution ............................................ Voltage: 0.1 V
Accuracy .............................................. Voltage: ±(1% +2 digits)
Stability at 100% load change .............. Voltage: <20 mV
Stability at 10% load change ................ Voltage: <4 mV
Residual ripple at nominal load ...........Voltage: <30 mVpp
Efficiency ..............................................>80%
Output power ........................................max. 81 W
Fuse type ..............................................Microfuse 5 x 20 mm (Ø x L), 250 V, 2 A, slow-acting tripping characteristics
Ambient conditions ............................... Temperature 0 °C to +40 °C, humidity max. 80% relative, non-condensing
Dimensions (W x H x D) .......................62 x 138.8 x 281.5 mm
Weight .................................................approx. 1 kg
28
Area 2: 0 - 27 V/DC
Area 3: 0 - 36 V/DC
Area 2: 0 - 3 A
Area 3: 0 - 2.2 A
Current: 0.01 A
Current: ±(1% +2 digits)
Current: <10 mA
Current: <10 mA