The Light Indicator Of The Front Panel Is Fixed But No Image Appears On The Screen; The Image Is Blurred; After-Sales Service And Maintenance; Technical Data - Delmont imaging iCare Guide D'utilisation

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

If the defect persists and it is necessary to return the device to the after-sales service, take care to send it to us
in its original packaging after having disinfected it.
8.4. The light indicator of the front panel is fixed but no image appears on the screen
- Check that the camera head is connected to the control unit (otherwise a colour bar will be displayed).
- Check that the control unit is correctly connected to the monitor (cable in good condition and plugs
properly inserted).
- Check that the monitor is turned on, that the correct video input is selected and that the screen image
settings are not in the minimum position (colour, light and contrast).
- Check for the presence of light by inspecting the light source, light cable and endoscope.
8.5.

The image is blurred

- Check that there is no fog on the camera lens or endoscope.
- Check the focus of the lens.

9. After-sales service and maintenance

No specific maintenance is required on this device.
Simply make sure that the aerators are not clogged with dust. If this is the case, unplug the unit and vacuum the
dust.
Errors of use are not covered by the warranty.
If a defect persists and it is necessary to return the device to the after-sales service, take care to ship it in its
original packaging. Also, the entire device (control unit, camera head, lens, cables, etc.) must be returned.
Please attach a small explanatory note to the shipment regarding the defect found.
The equipment must be disinfected before returning for repair.
When returning the equipment, check its condition and express reserves on the delivery order if necessary, confir-
ming them to the carrier by registered letter within 48 hours. In the case where a material dispatched by us would
suffer damage during its transport, the amount of the repair is charged either to the carrier if the reserves were
made within the time limits, or to the recipient in the contrary case.
In case of an incident, please contact our customer service or our nearest sales representative.

10. Technical data

Camera side:
- Waterproof HD CMOS sensor (IP67)
- 22 mm C-mount lens
- Resolution: 1920 x 1080
- Interlaced scanning
- Definition > 900 lines
- Sensibility : 2 000 lux at F8
- Signal-to-noise ratio: 54 dB
- Automatic electronic shutter (1/50 to 1/10 000)
D900 700 053 B
Instructions for use: iCare imaging station
Rev 09/2018
delmont
imaging
11

Publicité

Table des Matières
loading

Table des Matières