5.1 SELECT TUBING AND PUMP
HEADS
Use Table 1 (See below) to select the tubing to provide the
required single channel flow rate.
See Appendix B (Page 8) for ordering information on the
MASTERFLEX Two-Stop Pump Tube Sets.
5.2 SELECT PUMP SPEED
Select fixed speed drive or adjust speed of variable-speed
drive to provide desired nominal flow rate within the rpm
speed range shown for the pump model selected.
Pump
Models
07534-04
07534-08
07535-04
07535-08
07536-02
07536-04
* Not recommended to use these tubing sizes at speeds greater than 300 rpm.
6
Tubing
Type
0.19 mm ID
0.25 mm ID
0.89 mm ID
Microbore
1.42 mm ID
2.06 mm ID
2.79 mm ID
®
L/S
Precision
®
L /S
High
Performance
Precision
6. MAINTENANCE
No maintenance is required. Wipe the Pump Head with a
clean cloth and a mild detergent to clean. Never immerse or
use excessive fluid.
6.1 SERVICE PARTS
Service Kit
Consists of:
Mounting Screws (4)
Hex Key (1)
Support Foot (1)
Table 1 — Flow rates
Tubing
Drive RPM
Size
1 to 100
6 to 600
1 to 100
6 to 600
1 to 100
6 to 600
1 to 100
6 to 600
1 to 100
6 to 600
1 to 100
6 to 600
1 to 100
®
L/S
13
6 to 600
1 to 100
®
L/S
14
6 to 600
1 to 100
®
L/S
16
6 to 600
1 to 100
®
L/S
15
6 to 600
1 to 100
®
L/S
24
6 to 600
1 to 100
®
L/S
35
6 to 600
07534-01
Flow Rate per
Channel in ml/min
0.0026 to .26
Not recommended*
0.0053 to .53
Not recommended*
0.054 to 5.4
0.30 to 30
0.12 to 12
0.67 to 67
0.23 to 23
1.25 to 125
0.35 to 35
2.0 to 200
0.045 to 4.5
0.27 to 27
0.16 to 16
0.96 to 96
0.47 to 47
2.8 to 280
1.6 to 160
10 to 1000
2.8 to 280
17 to 1700
3.8 to 380
23 to 2300