Trouble-shooting
Symptom
The shower table
does not stay in
upright position .
The product will not
work, even though
the remote control is
used correctly .
The shower table
runs up and down
unevenly .
Contact Pressalit or your dealer if the problem cannot be remedied . Do not
attempt to repair the product yourself, as this will invalidate the guarantee .
Cause
The gas cylinders is
defective .
Fault in power supply .
The sliding brackets
are not properly
lubricated .
Action
Contact Pressalit or your
local dealer .
Check that all wires and
plugs are intact and
correctly connected .
Contact Pressalit or your
local dealer .
Oil the sliding brackets
as described under
"Maintenance" .