REPROCESSING IN A HOME
ENVIRONMENT
The product components must be cleaned thoroughly immedi-
ately after each application and disinfected at least once a
week.
The connection tubing cannot be cleaned or disinfected.
Dry the connection tubing after each use [see: Care of the con-
nection tube, page 27].
The maximum operating life of the connection tubing is 1 year.
4.1 Reprocessing cycles
Nebuliser
4.2 Processing limits
Nebuliser and accessories,
disinfection
4.3 Preparation
• Remove the tube from the nebuliser.
• Pull the mouthpiece off the nebuliser.
• Make sure that all residual volume is removed from the
nebuliser.
• Dismantle the nebuliser into all individual parts.
• Carefully pull the blue exhala-
tion valve out of the slot in the
mouthpiece. The exhalation
valve must still be attached to
the mouthpiece.
– 22 –
– Clean immediately after every use
– Disinfect once a week
300 processing cycles,
max. 1 year