EMC immunity influence ..................................... < ±0.5% of span
Extended EMC immunity:
NAMUR NE 21, A criterion, burst ....................... < ±1% of span
Auxiliary supplies:
2-wire supply (terminal 44...43) ................... 25...16 VDC / 0...20 mA
Max. wire size .............................................. 1 x 2.5 mm
Screw terminal torque ................................. 0.5 Nm
Relative humidity ......................................... < 95% RH (non-cond.)
Dimensions, without display front (HxBxD) .. 109 x 23.5 x 104 mm
Dimensions, with display front (HxBxD) ...... 109 x 23.5 x 116 mm
Protection degree (enclosure / terminals) ... IP50 / IP20
Weight ......................................................... 170 g / 185 g with PGMMOD00
RTD, linear resistance and potentiometer input:
Input
Min.
type
value
Pt100
-200°C
+850°C
Ni100
-60°C
+250°C
0 Ω
10000 Ω
Lin. R
10 Ω
Potentiometer
100 kΩ
Input for RTD types:
Pt10, Pt20, Pt50, Pt100, Pt200, PT250, Pt300, Pt400, Pt500, Pt1000
Ni50, Ni100, Ni120, Ni1000
Cable resistance per wire (max.), RTD ........ 50 Ω
Sensor current, RTD .................................... Nom. 0.2 mA
Effect of sensor cable resistance
(3- / 4-wire), RTD ......................................... < 0.002 Ω / Ω
Sensor error detection, RTD ....................... Yes
Short circuit detection, RTD ........................ < 15 Ω
10
10
All manuals and user guides at all-guides.com
2
stranded wire
Max.
value
Standard
IEC60751
DIN 43760
-
-
TC input:
Min.
Type
value
B
+400°C
+1820°C
E
-100°C
+1000°C
J
-100°C
+1200°C
K
-180°C
+1372°C
L
-200°C
+900°C
N
-180°C
+1300°C
R
-50°C
+1760°C
S
-50°C
+1760°C
T
-200°C
+400°C
U
-200°C
+600°C
W3
0°C
+2300°C
W5
0°C
+2300°C
LR
-200°C
+800°C
Cold junction compensation (CJC)
via internally mounted sensor ................. < ±1,0 °C
Sensor error detection, all TC types ........... Yes
Sensor error current:
when detecting ....................................... Nom. 2 µA
else .......................................................... 0 µA
Current input:
Measurement range .................................... -1...25 mA
Programmable measurement ranges .......... 0...20 and 4...20 mA
Input resistance ........................................... Nom. 20 Ω + PTC 50 Ω
Sensor error detection:
Loop break 4...20 mA ............................. Yes
Voltage input:
Measurement range .................................... -20 mV...12 VDC
Programmable measurement ranges .......... 0...1 / 0.2...1 / 0...5 / 1...5 /
Input resistance ........................................... Nom. 10 MΩ
Current output:
Signal range (span) ...................................... 0...20 mA
Programmable signal ranges ...................... 0...20 / 4...20 /
Load (max.) .................................................. 20 mA / 800 Ω / 16 VDC
Load stability ............................................... ≤ 0.01% of span / 100 Ω
Sensor error detection ................................ 0 / 3.5 / 23 mA / none
NAMUR NE 43 Upscale / Downscale ......... 23 mA / 3.5 mA
Max.
value
Standard
IEC 60584-1
IEC 60584-1
IEC 60584-1
IEC 60584-1
DIN 43710
IEC 60584-1
IEC 60584-1
IEC 60584-1
IEC 60584-1
DIN 43710
ASTM E988-90
ASTM E988-90
GOST 3044-84
0...10 and 2...10 VDC
20...0 and 20...4 mA
11
11