Immunity to proximity fields from RF wireless communication
equipment
Test
standard
EN 61000-4-3*
en
Electrical fast transient/burst
Test
standard
EN 61000-4-4
32
The product must not be affected by radiating RF signals,
for example from a mobile phone or radio, that could
cause the product to take action or to make components
defective .
Immunity test level
385 MHz, 27 V/m, 18 Hz PM (50 % duty cycle square wave)
450 MHz, 28 V/m, FM +/- 5 kHz deviation, 1 kHz sine
710 MHz, 9 V/m, 217 Hz PM (50 % duty cycle square wave)
745 MHz, 9 V/m, 217 Hz PM (50 % duty cycle square wave)
780 MHz, 9 V/m, 217 Hz PM (50 % duty cycle square wave)
810 MHz, 28 V/m, 18 Hz PM (50 % duty cycle square wave)
870 MHz, 28 V/m, 18 Hz PM (50 % duty cycle square wave)
930 MHz, 28 V/m, 18 Hz PM (50 % duty cycle square wave)
1720 MHz, 28 V/m, 217 Hz PM (50 % duty cycle square wave)
1845 MHz, 28 V/m, 217 Hz PM (50 % duty cycle square wave)
1970 MHz, 28 V/m, 217 Hz PM (50 % duty cycle square wave)
2450 MHz, 28 V/m, 217 Hz PM (50 % duty cycle square wave)
5240 MHz, 9 V/m, 217 Hz PM (50 % duty cycle square wave)
5500 MHz, 9 V/m, 217 Hz PM (50 % duty cycle square wave)
5785 MHz, 9 V/m, 217 Hz PM (50 % duty cycle square wave)
The product must not be affected by rapid pulse signals
through the cord . This could cause the components to
become defective, or the product could initiate actions .
Immunity test level
+/- 1 kV , +/- 2 kV
100 kHz repetition frequency