Electromagnetic compatibility
The device is suitable for use at home, at daycare centres, at day centres
for persons with physical or mental disabilities or at hospitals except near
active HF surgical equipment and RF shielded rooms with medical systems
for magnetic resonance imaging with high intensity of electromagnetic
disturbances .
The device does not have any function related to basic safety or essential
performance . In the event of performance loss or degradation caused by
electromagnetic disturbances, the user's safety remains uncompromised .
Cable
Power cable
Hand control spiral cable
en
• Use of accessories, transducers and cables other than those specified or
provided by the manufacturer could result in increased electromagnetic
emissions or decreased electromagnetic immunity of this equipment
and result in improper function .
• Portable RF communications equipment (including peripherals such as
antenna cables and external antennas) should be used no closer than
30 cm (12 inches) to any part of the unit, including its cables, to avoid
any degradation of the performance of this unit .
• Use of this equipment adjacent to or stacked with other equipment
should be voided because it could result in improper operation . If such
use is necessary, this equipment and the other equipment should be
observed to verify that they are operating normally .
30
Length
3,2 m (126")
Approx . 5 m (197") extended
WARNING
!