Troubleshooting

Fault
Display is off and no
signals from transmitter
Display is on but no signals
from transmitter
Display is off but there are
signals from transmitter
Display does not show the
correct data

10. Cleaning, maintenance and repair

Cleaning
Clean, disinfect or sterilize sensor as needed
f
(CIP/SIP).
Maintenance
Regular maintenance is not required.

11. Disposal

Do not dispose of in household
f
waste.
Separate materials and dispose
f
of in compliance with nationally
applicable regulations.

12. Accessories

For adapter and other accessories see www.baumer.com.
www.baumer.com
Cause
Sensor not correctly connected
Device error
Short circuit
UnitCom cable not connected
Unsuitable media characteristics
Operating Instructions
Conductivity measurement
CombiLyz AFI4/AFI5
Inductive conductivity transmitter
Action
Check plug and power supply.
f
Dismount and return sensor.
f
Remedy short circuit.
f
Connect the UnitCom cable between
f
display and transmitter.
Check signal quality with
f
FlexProgrammer.
Repair
Do not repair the sensor yourself.
Send damaged sensor to Baumer.
f
13 / 52