Select conductivity range.
f
Menu
AFIx menu
If desired, select or define further settings:
f
Concentration output
Display colors
Warnings
Relays
7.2 Configuration via FlexProgram
9 FlexProgrammer is connected.
Possibilities with the AFIx transmitter:
Configure temperature compensation
Select temperature source for compensation
Select conductivity range
Set current limits
Select concentration output
Configure media screen
Make a data logging
Calibrate sensor
Calibrate media
Possibilities with the DFON display:
Select screen layout
Select backlight intensity
Define relays
Define warning and error indication
Calibrating the sensor (conductivity/temperature)
9 AFIx is powered up
9 The used medium has a known conductivity/temperature
Adjust an offset for the conductivity/temperature for the sensor.
f
If necessary, reset the sensor offset in the FlexProgram.
f
Calibrating a medium
For the different ranges it is possible to calibrate the medium by calculating the temperature compensation
after 3 measurements.
9 Calibration is done in a controlled environment
9 AFIx is powered up
Have the sensor tip immersed in the medium for 1 minute before measuring the conductivity.
f
Measure the conductivity at 3 different temperatures using the same medium.
f
Calculate the temperature compensation for the medium.
f
For further information see HELP section in FlexProgram.
www.baumer.com
Input con g.
Conductivity ranges
Operating Instructions
Conductivity measurement
CombiLyz AFI4/AFI5
Inductive conductivity transmitter
Conductivity range 1
Conductivity range 2
Conductivity range 3
Conductivity range 4
Select range
Select range
Select range
Select range
11 / 52