Troubleshooting
a) The output value from the sensor is higher than the value detected
with DPD:
The porous membrane is damaged
Agents water pollutants interfere with the measurement
Wiring damaged
pH of the water is lower than 6
b) The output value from the sensor is lower than the value detected
with DPD:
Are there any deposits on the porous membrane
The working flow is too low
There are air bubbles on the porous membrane
The presence of surfactants in water
pH of the water is higher than 8
There is more electrolyte inside the membrane cap
The gold electrode is oxidized
c) The output value from the sensor is 4mA (0ppm)
In water is present only combined chlorine
Content of free chlorine to below the limit of detection
The sensor is not properly connected
The sensor is faulty
d) The output value from the sensor is unstable
There are air bubbles on the membrane
The porous membrane is damaged
The problem does not depend on the sensor