Télécharger Imprimer la page

IFM Electronic efector500 PI76 Notice Utilisateurs page 17

Capteur de pression électronique

Publicité

Les langues disponibles

Les langues disponibles

1. Function and features
• The pressure sensor detects the system pressure,
• shows the current system pressure on its display (indication as
from 1% to 105% of the value of the measuring range),
• and generates 2 output signals according to the set output
configuration.
Switching function
(can be selected for each
output separately)
The unit is EHEDG 3A certified
Applications
Type of pressure: relative pressure
Order no.
PI76x2
PI76x3
PI76x4
PI76x6
PI76x7
PI76x8
Avoid static and dynamic overpressure exceeding the given over-
load pressure.
For gaseous media the application is limited to max. 25 bar.
Even if the bursting pressure is exceeded only for a short time
the unit can be destroyed (danger of injuries)!
Indication of the current system pressure as from 1% of the
value of the measuring range. Display "0" does not mean that
the system is free of pressure!
Measuring range
0 ... 100 bar
0 ... 25 bar
0 ... 10 bar
0 ... 2.5 bar
0 ... 1 bar
0 ... 250 mbar
output 1
hysteresis function / N.O. (Hno)
hysteresis function / N.C. (Hnc)
window function / N.O. (Fno)
window function / N.C. (Fnc)
Permissible
overl. pressure
200 bar
100 bar
50 bar
20 bar
10 bar
10 bar
output 2
Bursting
pressure
650 bar
350 bar
150 bar
50 bar
30 bar
30 bar
17