A. Spécifications techniques
Guidance and manufacturer's declaration – electromagnetic immunity
The ICR-1 is intended for use in the electromagnetic environment specified below. The
customer or the user of the ICR-1 should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Electrostatic
6 kV contact
discharge (ESD)
IEC 61000-4-2
8 kV air
Electrical fast
2 kV for power supply
transients/bursts
lines
IEC 61000-4-4
1 kV for input/output
lines
Surge
1 kV differential mode
IEC 61000-4-5
Voltage dips,
<5 % U
short
(>95 % dip in U
interruptions and
for 0.5 cycle
voltage variations
on power supply
40 % U
lines
(60 % dip in U
IEC 61000-4-11
for 5 cycles
70 % U
(30 % dip in U
for 25 cycles
<5 % U
(>95 % dip in U
for 5 sec
Power frequency
3 A/m
(50/60 Hz)
magnetic field
IEC 61000-4-8
NOTE U
is the a.c. mains voltage prior to application of the test level.
T
Manuel d'utilisation 204878
Compliance level
6 kV contact
8 kV air
2 kV for power
supply lines
1 kV for
input/output lines
1 kV differential
mode
<5 % U
T
)
(>95 % dip in U
T
for 0.5 cycle
40 % U
T
)
(60 % dip in U
T
for 5 cycles
70 % U
T
)
(30 % dip in U
T
for 25 cycles
<5 % U
T
)
(>95 % dip in U
T
for 5 sec
3 A/m
Appareil d'imagerie
Electromagnetic
environment - guidance
Floors should be wood,
concrete or ceramic tile.
If floors are covered with
synthetic material, the
relative humidity should
be at least 30 %.
Mains power quality
should be that of a
typical commercial or
hospital environment.
Mains power quality
should be that of a
typical commercial or
hospital environment.
Mains power quality
T
)
should be that of a
T
typical commercial or
hospital environment. If
user of the ICR-1
T
)
requires continued
T
operation during power
mains interruptions, it is
recommended that the
T
)
ICR-1 be powered from
T
an uninterruptible power
supply or a battery.
T
)
T
Power frequency
magnetic field should be
at levels characteristic
of a typical location in a
typical commercial or
hospital environment.
A - 5