Télécharger Imprimer la page

R82 High-low Mode D'emploi page 31

Masquer les pouces Voir aussi pour High-low:

Publicité

Service information
Maintenance of a medical device is wholly the responsibility of the owner of that device. Failure to maintain a
device in accordance with the instructions may invalidate the device's warranty. Furthermore, failure to maintain
a device may compromise the clinical condition or safety of users and/or their caregivers.
1st Day
Read the User Guide carefully before use and
store it in a safe place for future reference
Store the enclosed tools on/at the product
Daily
Use a dry cloth to clean the product
Visual inspection. Repair or replace any
damaged or worn-out parts
Ensure that all fasteners are in place and fitted
correctly
Check all fittings and buckles for signs of worn
out parts
Be aware of any MAX mark indications before
adjustment
Check that all cables connecting the power
system are in good shape
Weekly
Check that all wheels are moving freely and
that all wheel locks/brakes are easily activated
Wipe off any residue and dirt from the product
using a cloth with warm water and a mild
detergent/soap and dry before use
Monthly
Check the tire pressure: Front wheels 35
PSI/2.4 bar; Rear wheels 40 PSI/ 2.75 bar
All nuts and bolts on this product must be
checked and tightened to avoid unnecessary
failures
Oil the swing-away parts. We recommend you
use a professional lubricant system
Annually
Inspect the frame for cracks or signs for worn-
out parts and carry out annual service. Never
use a product which appears to be substandard
or faulty
Warewashing machine
This product can be washed at 60° using a mild
detergent in 15-20 minutes in a washing machine
designed for medical devices. Use the machine's
drying function to dry the product. Clean the
accessories seperately.
Before washing
Avoid degreasing of the piston rods, the
actuators should be retracted to minimum
stroke and without load before washing.
Remove the following items and clean them
separately: Seat, Accessories, Power Unit and
Handset
After washing:
Install power unit and handset
Check that the handset functionalities works
correct
support.R82.org
All manuals and user guides at all-guides.com
Disinfection
The product can be disinfected with a 70%
disinfectant IPA solution. It is recommended to wipe
off any residue and dirt from the product, using a
cloth with warm water and a mild detergent/soap
without chlorine and let it dry before disinfection.
Service interval
A detailed inspection of the product must be made
every 12 months (every 6 months for heavy users)
and every time the product is reissued for use. The
inspection should be carried out by a technically
competent person who has been trained in the use
of the product.
The lifetime of this product in normal use is 5 years
if all maintenance and servicing is carried out in
accordance with the manufacturer's instructions
and demonstrably recorded.
Service Check List:
Check and adjust the following functions as per their
intended use. Repair or replace any damaged or
worn-out parts. Please check the:
Frame is adjustable in height
Powered versions are able to retain power after
charging the battery
On outdoor versions, the push-brace can be
folded over the frame
Tire pressure
Seat plane adjustment. The seat plane
is adjustable and stays in position after
adjustment.
Wheels are moving freely and that all wheel
locks/brakes are easily activated. Adjust if
required
Nuts and bolts on the product are tightened
Oil the swing-away parts
Product for cracks or signs for worn-out parts
All labels on the product are intact
31
EN

Publicité

loading

Ce manuel est également adapté pour:

High-low:x