Electrical
fast
transients /
61000-4-4
bursts
Surges
Line-to-line
61000-4-5
Surges
Line-to-ground
61000-4-5
Conducted
disturbances
induced by RF
61000-4-6
fields
RATED
power
frequency
61000-4-8
magnetic fields
Voltage dips
61000-4-11
Voltage
interruptions
61000-4-11
Mesures
• Système de mesure par caméra vidéo et processeur de traitement embarqué
• Gamme de mesure : 1 à 9 mm (Diamètre de la pupille)
• Précision 0,1 mm ou 5%, résolution 0,01 mm (Taille de la pupille)
• Fréquence des mesures et de l'acquisition image 60 image/s.
Transfert des données
• Transfert des données par Bluetooth (selon système de recueil)
- Bande de fréquence: [2400 - 2483,5] MHz
- Puissance de sortie maximum: 8 dBm
Stimulation lumineuse de la pupille
• Durée de 1 seconde
• Intensité fixe à 320 Lux (précision +/-10%)
Manuel Utilisateur NeuroLight
± 2 kV
IEC
100 kHz
repetition
frequency
IEC
± 0,5 kV, ± 1 kV
± 0,5 kV, ±
IEC
1 kV,
± 2 kV
3 V
0,15 MHz –
80 MHz
6 V in ISM
IEC
bands
between 0,15
MHz and
80 MHz
80 % AM at
1 kHz
IEC
30 A/m
50 Hz or 60 Hz
0 % UT;
0,5 cycle
At 0°, 45°, 90°,
135°, 180°, 225°,
270° and 315°
IEC
0 % UT;
1 cycle and
70 % UT; 25/30
cycles
Single phase:
at 0°
IEC
The NeuroLight may temporarily not
± 2 kV
display
100 kHz
electromagnetic disturbances such
repetition
as the use of electrosurgery device. In
frequency
that case, the NeuroLight will maintain
the safety of the patient and the user.
Mains power quality should be that of
± 0,5 kV, ± 1 kV
a typical residential, commercial or
hospital environment.
Mains power quality should be that of
± 0,5 kV, ± 1
a typical residential, commercial or
kV, ± 2 kV
hospital environment.
3 V
0,15 MHz –
80 MHz
6 V in ISM
In order to prevent electromagnetic
bands
disturbance,
between 0,15
separation from RF communication
MHz and
equipment of 30cm
80 MHz
80 % AM at
1 kHz
Mains power quality should be that of
30 A/m
a typical residential, commercial or
50 Hz or 60 Hz
hospital environment
0 % UT;
0,5 cycle
At 0°, 45°, 90°,
135°, 180°, 225°,
270° and 315°
Mains power quality should be that of
a typical residential, commercial or
0 % UT;
hospital environment.
1 cycle and
70 % UT; 25/30
cycles
Single phase:
at 0°
Mains power quality should be that of
a typical residential, commercial or
hospital environment.
result
during
transient
keep
minimum
22