C3.7 Confirming the HLD Cycle
1.
Close the trophon2 chamber door and follow the onscreen instructions to record the Chemical
Indicator result.
2.
The trophon2 device will indicate a passed or failed cycle. If the trophon Chemical Indicator and/or
trophon2 device indicate a failed cycle, repeat steps from Section C3.1 Preparing the Probe
A disinfection cycle will only be completed and logged once the operator has confirmed
the Chemical Indicator result, and scanned their Operator Card to confirm (if AcuTrace
AND Log Operator Card is enabled). Any errors or power-related issues should be
considered a fail. Refer to SECTION F1: Incomplete or Failed Cycles.
3.
If the trophon2 device indicates a passed cycle, the probe is now ready for use or storage in a
clean, single-use, approved probe storage solution such as the Clean Ultrasound Probe Cover.
Once stored, discard the gloves.
The medical instrument (probe) is ready for patient use only after the disinfection cycle is completed as
described above.
C3.8 Sleep Mode
trophon2 has a power-saving mode to reduce power when inactive. Sleep mode will be entered either:
By inactivity for an extended period of time (default two hours). This can be adjusted or turned
off through selecting Menu
At a particular time of day using a daily timer. This can be adjusted by selecting Menu
Settings
Daily timer.
Sleep mode will be indicated by the onscreen symbol:
Touch the symbol to wake up the trophon2 device to prepare it for use.
Settings
Sleep timer.
27
L03082 4.0 01/2023