TROUBLESHOOTING
All manuals and user guides at all-guides.com
If any abnormality arises during use, please check the following
SYMPTOMS
Unit does not
turn on when the
START/STOP
button is pushed.
EE measurement
error symbol shown
on display or the
blood pressure
value is displayed
excessively low (or
high).
E1 error symbol
shown on display
E2 error symbol
shown on display
E3 error symbol
shown on display
points:
Note: If the unit still does not work, contact HoMedics Consumer
Relations. Under no circumstance should you disassemble
or attempt to repair the unit by yourself. Contact informa-
tion for HoMedics Group Canada Department can be
found on the warranty page (See page 34).
POSSIBLE
CAUSES
Batteries have run
down.
Battery polarities
have been positioned
incorrectly.
Cuff has been placed
incorrectly.
Did you talk or
move during
measurement?
Shaking of the wrist
with the cuff on.
Air circuit
abnormality.
Inflation pressure
exceeding 300 mmHg.
Error determining
measurement data.
32
CORRECTION
Replace them
with two new AAA
alkaline batteries.
Re-insert the
batteries in the
correct positions.
Wrap the cuff
proerly so that it is
positioned correctly.
Measure again.
Keep wrist
steady during
measurement.
Measure again.
Switch the unit off,
then measure again.
Measure again.