Multifunction Installation Tester/English
10.4-Drawing
No.
Annunctator
File Name
Primary display and measurement units.
Primary display and measurement units.
Coordinate
Function
Function
hours/minutes/seconds
L-FE Value
L-N Value
FE-N Value
Arrows above or below the terminal
indicator symbol indicate reversed polarity.
Check the connection or check the wiring
to correct.
UF Value
Main display curve
34
Meaning
File named:
Month/day File type
Function
Hours/minutes/seconds
Record time
10.5-Datalog Color
WARNING!
• Measurements should only be performed on de-energized circuit
• Measurements may be adversely a ected by impedances or paral
To Measure Continuity
1-Turn the rotary switch to the RLO position.
2-Use the L and N (red and black) terminals for this test.
3-Before making a continuity test, short the ends of the probes
performing test leads compensation compensated test leads indicator is displayed.
4-Press and hold TEST until the reading settles.
5-If the continuity beeper is enabled,press the F1 to set high
for measured values less than high limit resistance and there is no stable reading beep for measured values
greater than high limit resistance.
Multifunction Installation Tester/English
s.
lel circuits or transient currents.
together and press the ZERO button. After
limit resistance value. the tester beeps continuously
35