5. Mode Control
Expose the controls by sliding front cover down. This switch has 3
positions: B for Burst stimulation, N for Constant stimulation, and M
for modulated stimulation. Push the Mode Selector to the desired
position.
6.Pulse Rate Control:
This control determines the frequency (pulses per second). This dial
controls the frequency to both channels. Unless otherwise instructed
set the pulse rate to between 70 - 120 Hz.
7.Pulse Width Control:
This control adjusts the length of each electrical impulse. By
increasing the pulse width the strength and sensation of the
stimulus will be increased. By reducing the pulse width the strength
and sensation will be reduced. Unless otherwise instructed set the
control to between 70 - 120
.
µs