P. Changing the device fuse
The device fuse is located on the bottom of the housing. Before replacing the fuse, please unplug the
power cord. To move with a screwdriver, unscrew the fuse holder and micro fuse T2.5 A (for 115 volt
version: replace T5.0 A).
In case of malfunction:
Check for the correct power supply.
Ensure that the hand piece is connected to the device properly.
Check, if the dust bag is torn.
Check, if the handle is assembled correctly.
When using a foot pedal, check if the device works without the pedal.
The problem is still not corrected; please do not hesitate to contact your service centre immediately.
S
PECIAL QUALITY FEATURES
Keypad:
Extremely easy to use, select all functions by pressing the corresponding key on the keypad.
LED display:
On the large, easy-to-read display, you are provided with all information about the device at a glance.
Quick chucks:
For the absorption of the abrasives, the quick chuck enables you a simple, fast and smooth change of the
grinders. Once the abrasive body begins to rotate, the shaft is automatically locked.
Overload protection for motor hand piece
The unit is equipped with overload protection for the hand piece. In case of any blockage or overload of the
hand piece, the engine speed as well as the vacum turbine is turned off for about 5 seconds.
Thermal cutoff of the suction turbine
If the suction turbine is not cooled adequately and the air temperature in the suction turbine housing exceeds
80° C, the suction turbine and the handpiece are switched off.
Control system for recommended dust bag replacement:
This dust bag control system indicates the remaining operating time with the dust bag used and can be
adjusted individually by the user based on personal experience.
M
AINTENANCE TIPS
To ensure that the safety, reliability and performance of the unit are not adversely affected, enhancements,
readjustments, alterations or repairs are not permitted aside from those performed by Promed authorized
persons.
A. Keep the inner parts of the handpiece (Collet chuck/quick lock chuck) as free as possible from dust and
grinding residue.
B. Remove dirt from the outer surface of the handpiece with a cloth moistened with alcohol. Prevent moisture
from entering the unit.
C. Only use grinding heads with a dental standard size (ø 2.332 – 2.350 mm (0.091 - 0.093 inches), shaft
length 28 - 36 mm) and avoid using rusted or damaged grinding heads. Their false usage may lead to
excessive wear and tear in the ball bearings. Adhere to the safe speed and the safety measures stipulated
by the manufacturer.